| Model Number |
X412H |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/30/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that a patient underwent a pacemaker placement on (b)(6) 2021 and suture was used.While suturing down the pacer internally in the patients left upper chest area the surgeon did the first suture pass and grabbed the needle with the drivers and placed it on the field just outside the incision.When the surgeon went to pass the suture needle a second time, the surgeon realized the tip was broken as it would not pass.The procedure was stopped and the team looked and x-rayed and magnified in on the pocket and pacer looking for the tip.Nothing could be seen under fluoroscopy but then found the tip on the sterile field.There was maybe a 5 minute delay without any negative outcome to the patient.Additional information was requested.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Additional information was requested, and the following was obtained: was there any additional tissue damage as a result of searching for the needle piece? no.Was there any change in the patient¿s post-operative care due to the prolonged procedure? unspecified.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4) date sent to the fda: 3/3/2022 additional information: h3, h6 h3 investigational summary: a picture received to ethicon inc.For evaluation.As per visual inspection of the picture a needle of product code x412 was compared against the drawing printed in paper lid and it was observed that the tip needle was damaged or broken, body fluids could be noted on the strand and damaged near of the swage area of the needle.As part of our quality process, the manufacturing records of this lot-serial number were reviewed, and the manufacturing standards were met prior to the release of this batch.No conclusion could be reached as to what caused the reported complaint since the sample was not returned for analysis.As part of ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint (b)(4).Date sent to the fda: 02/23/2022.H6 component code: g07002 - device not returned.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A manufacturing record evaluation was performed for the finished device, and no non-conformances were identified.
|
| |
|