• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. T-HANDLE; APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. T-HANDLE; APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT Back to Search Results
Model Number 71631172
Device Problem Crack (1135)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/06/2022
Event Type  malfunction  
Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event Description
It was reported that, during a nail procedure, the t-handle cracked outside the patient.No pieces fell into patient.The procedure was completed, without delay, using the same device.Patient was not harmed as a consequence of this issue.
 
Manufacturer Narrative
H3, h6: the device was not returned for evaluation and the reported event could not be confirmed.A review of complaint history for the previous 12 months did not reveal similar events for the listed device.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.The contribution of the device to the reported event could not be corroborated.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
T-HANDLE
Type of Device
APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key13353157
MDR Text Key284476234
Report Number1020279-2022-00313
Device Sequence Number1
Product Code KTT
UDI-Device Identifier03596010425096
UDI-Public03596010425096
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K093047
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71631172
Device Catalogue Number71631172
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/06/2022
Initial Date FDA Received01/25/2022
Supplement Dates Manufacturer Received02/25/2022
Supplement Dates FDA Received02/28/2022
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-