Model Number AR40E
Medical Device Problem Codes Break (1069); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Codes Capsular Bag Tear (2639); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 11/25/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that during surgery and prior to intraocular lens (iol) implantation, the posterior capsule ruptured.The doctor proceed to implant the intraocular lens in the ciliary sulcus.The iol position was found to have been shifted after the surgery.Therefore, the patient was given intraocular lens internal fixation.During the secondary surgery, it as found that the intraocular lens haptic was transferred out from the corneal incision.The haptic was found to be broken.The corneal incision was enlarged and the intraocular lens was completely removed.A replacement lens of another brand was used.No further information available.
 
Additional Manufacturer Narrative
(b)(6).The device was not returned for evaluation; therefore, a failure analysis of the complaint device cannot be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.(b)(4).All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
Additional information received from customer on 21 feb 2022 indicated that a posterior capsule tear occurred during lens implantation.No intervention was performed.Therefore, the information is being provided in this supplemental filing and the following section has been updated accordingly: section h6: health effect - clinical code: 2639.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENSAR IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key13354609
Report Number3012236936-2022-00158
Device Sequence Number1647911
Product Code HQL
UDI-Device Identifier05050474502178
UDI-Public(01)05050474502178(17)240219
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 03/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/19/2024
Device Model NumberAR40E
Device Catalogue NumberAR40E00235
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/04/2022
Supplement Date Received by Manufacturer02/21/2022
Initial Report FDA Received Date01/26/2022
Supplement Report FDA Received Date03/09/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/19/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age71 YR
Patient SexMale
-
-