Updated fields: d4, d9, g3, g6, h2, h3, h4, h6, h10.The microsensor kit (id 826638) was returned for evaluation.Failure analysis - the issue of the complaint was not confirmed.No visible damage to the millar sensor, catheter material, or connector.Icp express passed with reading 484.The device passed electronic, noise, linearity/hysteresis, and signal drift tests.The root cause of the issue reported by customer could not be determined as the device worked correctly.However, the possible root cause of the defect reported by the customer could be due to incorrect set-up of device.
|