ALLERGAN (COSTA RICA) STYLE 410 COHESIVE SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Model Number MF-410255
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 12/29/2021
Type of Reportable Event Serious Injury
Event or Problem Description
Healthcare professional reported "rupture" and "patient concern with the product" against right device.The device remains implanted.
 
Additional Manufacturer Narrative
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
 
Additional Manufacturer Narrative
Additional, corrected and/or changed data: b.5, d.6b, d.9, g.2, h.3, h.6.
 
Event or Problem Description
The device has been explanted and replaced.
 
Event or Problem Description
Healthcare professional reported "rupture" and "patient concern with the product" against right device.The device has been explanted and replaced.
 
Additional Manufacturer Narrative
Device evaluation: visual analysis of the returned device identified: underweight, broken shell and missing piece of the shell.A microscopic analysis was performed which identify a sharp edge broken assessed as unidentified tear broken and missing piece of the shell.A dimension measurement in the shell was performed which identify the thickness within specification.Based on the device analysis the final assessment is: a broken sharp in the posterior side assessed as unidentified (tear) opening.Missing piece of the shell assessed as inconclusive.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STYLE 410 COHESIVE SILICONE GEL FILLED BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
Manufacturer (Section G)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
Manufacturer Contact
chris sprague
12331-a riata trace parkway
building 3
austin, TX 78727
8479375615
MDR Report Key13358303
Report Number9617229-2022-00272
Device Sequence Number1663060
Product Code FTR
UDI-Device Identifier10888628003446
UDI-Public10888628003446
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
P040046
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/02/2019
Device Model NumberMF-410255
Device Catalogue NumberMF-410255
Device Lot Number2678868
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/31/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 01/04/2022
Supplement Date Received by Manufacturer01/11/2022
02/04/2022
Initial Report FDA Received Date01/26/2022
Supplement Report FDA Received Date02/03/2022
03/03/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/02/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age51 YR
Patient SexFemale
Patient Weight61 KG
-
-