| Model Number |
MF-410255 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
12/29/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Healthcare professional reported "rupture" and "patient concern with the product" against right device.The device remains implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional, corrected and/or changed data: b.5, d.6b, d.9, g.2, h.3, h.6.
|
| |
|
Event or Problem Description
|
|
The device has been explanted and replaced.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported "rupture" and "patient concern with the product" against right device.The device has been explanted and replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: visual analysis of the returned device identified: underweight, broken shell and missing piece of the shell.A microscopic analysis was performed which identify a sharp edge broken assessed as unidentified tear broken and missing piece of the shell.A dimension measurement in the shell was performed which identify the thickness within specification.Based on the device analysis the final assessment is: a broken sharp in the posterior side assessed as unidentified (tear) opening.Missing piece of the shell assessed as inconclusive.
|
| |
|