• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NIHON KOHDEN CORPORATION GF-210RA; MULTI-GAS UNIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NIHON KOHDEN CORPORATION GF-210RA; MULTI-GAS UNIT Back to Search Results
Model Number GF-210RA
Device Problems Unable to Obtain Readings (1516); Output Problem (3005)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/28/2021
Event Type  malfunction  
Event Description
The customer reported that the multigas unit intermittently stops reading gases.There was no patient injury reported.
 
Manufacturer Narrative
According to the customer, the unit displays no errors, it just stops generating readings.The customer will send in the unit to be exchanged.Nihon kohden continues to investigate the reported event.Nihon kohden will submit a supplemental report in accordance with 21 cfr section 803.56 when additional information becomes available.
 
Event Description
The customer reported that the multigas unit intermittently stops reading gases.There was no patient injury reported.
 
Manufacturer Narrative
Details of complaint: the customer reported that the multigas unit intermittently stops reading gases.According to the customer, the unit displays no errors, it just stops geerating readings.The customer will send in the unit to be exchanged.There was no patient injury reported.Investigation summary: the root cause of the reported issue is attributable to the state of wear-components such as watertrap, connectors, sampling line, and/or device damage.Manufacturer references # (b)(4) follow up 001.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GF-210RA
Type of Device
MULTI-GAS UNIT
Manufacturer (Section D)
NIHON KOHDEN CORPORATION
attn: shama mooman
1-31-4 nishiochia
shinjuku-ku, tokyo 161-8 560
JA  161-8560
Manufacturer (Section G)
NIHON KOHDEN TOMIOKA CORPORATION
attn: shama mooman
1-1 tajino
tomioka city, gunma 370-2 314
JA   370-2314
Manufacturer Contact
shama mooman
safety mgmt dept, quality mgmt
seibu bldg 2, 4th floor 1-11-2
kusunokidai tokorozawa, saitama 359-8-580
JA   359-8580
MDR Report Key13359356
MDR Text Key284619854
Report Number8030229-2022-02400
Device Sequence Number1
Product Code CCK
UDI-Device Identifier04931921106891
UDI-Public04931921106891
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K110594
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 07/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/26/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberGF-210RA
Device Catalogue NumberGF-210RA
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/08/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/11/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-