| Model Number |
1.75 |
| Medical Device Problem Code |
Patient Data Problem (3197)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/28/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
There is currently a pending investigation.Nova is requesting additional information and further details will be provided in a supplemental report.
|
| |
|
Event or Problem Description
|
|
The customer reports that a statstrip meter scanned a patient armband incorrectly.When one patient was scanned a different patient's data appeared.After a rescan the correct patient's information appeared so the meter's serial number was not taken note of.There was no patient harm or intervention reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Nova initiated an investigation into this meter mis-scan.By requesting, the patient armbands in question.The customer sent sample armbands.And they were graded using a barcode verifier.The quality of the armbands was very low.And it was concluded, that the mis-scan was caused by the poor quality of the armbands.The customer is reviewing their printers and settings, but the nova meter performed as it should.
|
| |
|