Addition information: h10: the device was evaluated on site by a qualified baxter service technician.Visual inspection observed that the four screws holding the main body of the machine to the stand were loosened.The screws were tightened.Three photos and one video were inspected as well.One of the photos showed that out of the four inner screws holding the main body of the machine stand, two were partially unscrewed and made the man body loose on the stand.However, the video showed that the displacement of the body allowed by the unscrewed bolt was minimal and its extension wasn¿t able to cause the machine tilting over.The reported condition was verified.The cause could not be determined.A device history review revealed no issues that could have caused or contributed to the reported issue.Should additional relevant information become available, a supplemental report will be submitted.
|