BOSTON SCIENTIFIC CORPORATION HYDRATOME RX 44; UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES) Back to Search Results
Model Number M00583050
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/06/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported to boston scientific corporation that a hydratome rx 44 was used in the bile duct during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2022.During the procedure, the cutting wire of the hydratome rx 44 broke.It was reported that no part of the cutting wire detached and fell into the patient.The procedure was completed with another of the same device.There were no patient complications reported as a result of this event.
 
Additional Manufacturer Narrative
The returned hydratome rx 44 was analyzed, and a visual evaluation noted that the cutting wire was broken, kinked, and blackened.The device was observed under magnification and the cutting wire was blackened and was broken from the proximal pierced hole.The preloaded guidewire of the device was not returned.No other problems with the device were noted.The reported event was confirmed.The product analysis revealed that the cutting wire was broken, kinked, and blackened.These conditions could have been generated if there was contact between the device and the scope during energization or if the device exceeded the maximum of voltage during the procedure.Also, energizing the device prior to performing sphincterotomy can compromise the cutting wire's integrity, which can cause a premature cutting wire fatigue.Additionally kinking/bending the cutting wire can lead to a break in the cutting wire.It is most likely that procedural or anatomical factors encountered during the use of the device could have affected the device performance and its integrity.Based on all gathered information, the most probable root cause of this complaint is adverse event related to procedure.A review of the manufacturing documentation for this device revealed that no anomalies or deviations related to the event occurred during manufacturing.A labeling review was performed and, from the information available, this device was used per the instructions for use / product label.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HYDRATOME RX 44
Common Device Name
UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
2546 calle primera
alajuela
CS  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key13359754
Report Number3005099803-2022-00193
Device Sequence Number13864880
Product Code KNS
UDI-Device Identifier08714729755203
UDI-Public08714729755203
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K013153
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 01/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/01/2024
Device Model NumberM00583050
Device Catalogue Number8305
Device Lot Number0028315902
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/17/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/06/2022
Initial Report FDA Received Date01/26/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/02/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexFemale
Patient RaceWhite
-
-