| Model Number |
MT22949 |
| Medical Device Problem Code |
Unable to Obtain Readings (1516)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/19/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that error icon displayed occurred.No product or data was provided for evaluation.The allegation and a probable cause could not be determined.No injury or medical intervention was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4) b5: describe event or problem - additional d9: device availability - additional h2: additional information/device evaluation h3: device evaluated by manufacturer - additional h6: adverse event problem - additional.
|
| |
|
Event or Problem Description
|
|
The product was evaluated.An external visual inspection was performed and passed.Receiver charge and boot tests were performed and passed.Functional tests were performed and passed.An internal visual inspection was performed and passed.The receiver log was downloaded and reviewed finding error related to the complaint.The allegation was confirmed.The probable cause could not be determined.No injury or medical intervention was reported.
|
| |
|