MAUDE Adverse Event Report: SISTEMAS MEDICOS ALARIS, S.A. DE C.V. BD ALARIS¿ PUMP MODULE SMARTSITE¿ INFUSION SET; INTRAVASCULAR ADMINISTRATION SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SISTEMAS MEDICOS ALARIS, S.A. DE C.V. BD ALARIS¿ PUMP MODULE SMARTSITE¿ INFUSION SET; INTRAVASCULAR ADMINISTRATION SET Back to Search Results
Model Number 10015896
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/07/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Medical device expiration date: unknown.A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.Device manufacture date: unknown.
 
Event or Problem Description
It was reported that the bd alaris¿ pump module smartsite¿ infusion set was damaged.There was no report of patient impact.The following information was provided by the initial reporter: product breaking at multiple junctures.
 
Event or Problem Description
It was reported that the bd alaris¿ pump module smartsite¿ infusion set was damaged.There was no report of patient impact.The following information was provided by the initial reporter: product breaking at multiple junctures.
 
Additional Manufacturer Narrative
Medical device expiration date: unknown.A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.Device manufacture date: unknown.
 
Additional Manufacturer Narrative
The following fields were updated due to additional information: d.9.Device available for eval?: yes.D.9.Returned to manufacturer on: 1/19/2022.H.6.Investigation: a complaint of product breaking in multiple areas was received from the customer.A sample was returned for investigation.The tubing looked as it had been sliced, separating the drip chamber from the top of the set.The stopcock was separated at the male luer and the female luer.Solvent was not noted on either connection points.No other damages or defects were observed.A device history record review could not be performed on model 100115896 because the lot number was unknown.The root cause for the separation was due to an insufficient amount of solvent being added to the connection sites, making the set easily separable.H3 other text : see h.10.
 
Event or Problem Description
It was reported that the bd alaris¿ pump module smartsite¿ infusion set was damaged.There was no report of patient impact.The following information was provided by the initial reporter: product breaking at multiple junctures.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD ALARIS¿ PUMP MODULE SMARTSITE¿ INFUSION SET
Common Device Name
INTRAVASCULAR ADMINISTRATION SET
Manufacturer (Section D)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer (Section G)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key13360862
Report Number9616066-2022-00032
Device Sequence Number19251383
Product Code FPA
UDI-Device Identifier50885403234062
UDI-Public50885403234062
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
K931173
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number10015896
Device Catalogue Number10015896
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/19/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/10/2022
Supplement Date Received by Manufacturer02/03/2022
Initial Report FDA Received Date01/26/2022
Supplement Report FDA Received Date02/16/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-