Prior to patient contact, it was discovered that the stylet of the cook quick-core coaxial biopsy needle set was screwed too tight in the outer part of the needle and was unable to be taken apart.Another device was used to complete the procedure.As reported, the patient did not experience any adverse effects or require any additional procedures due to this occurrence.
|
|
Investigation ¿ evaluation: nanjing changde med.Supp.Co.(china) contacted cook on 13jan2022 about the cook quick-core coaxial biopsy needle set (qcs-18-15.0-20t, lot# 13847972).Prior to patient contact, it was discovered the stylet was screwed too tightly in the outer part of the dtn needle and was unable to be taken apart.The procedure was completed with a new device.As reported, the patient did not experience any adverse effects or require any additional procedures due to this occurrence.Reviews of documentation including the complaint history, device history record (dhr), quality control procedures, specifications, and instructions for use (ifu), as well as a visual inspection and functional test of the returned device, were conducted during the investigation.The product was returned to cook for evaluation.During evaluation there was no difficulty noted in removing the stylet.Measurements of the dtn needle and stylet were made and were determined to be within specification.Additionally, a document-based investigation evaluation was performed.A review of the device master record (dmr) concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.A review of the dhrs for the reported complaint device lot 13847972 and the related subassembly lot sa13716020 did not have any non-conformances.There are no other complaints on this lot.Based on the available information, cook has concluded that there is no evidence suggesting nonconforming product exists either in house or in field.Cook also reviewed product labeling.This product is supplied with an instructions for use (ifu) pamphlet, t_qc_rev6.In the how supplied section it states: upon removal from package, inspect the product to ensure no damage has occurred.Based on the device master record, the device history record, and the device failure analysis, cook found no evidence the product was manufactured out of specification.There is no evidence of non-conforming material in house or in the field.Based on the information provided, inspection of the returned device, and the results of the investigation, cook concluded the cause for this failure cannot be established.During evaluation, cook was not able to reproduce the failure mode.There was no difficulty noted in removing the stylet.Measurements of the dtn needle and stylet were made and were determined to be within specification.The appropriate personnel have been notified.Cook will continue to monitor for similar complaints.Per the risk assessment no further action is required.Appropriate measures have been taken to address this failure mode.A capa was previously opened to further investigate this failure mode.Corrective actions have since been implemented following the manufacture of this device this report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|