It was reported that while preparing for a navio assisted tka surgery, the navio surgical system us booted up as expected but, when the attempt was made to home the bur they received a message saying "internal cabling/drill console error".After they rebooted, they got an error message came up saying "pfs control hardware failure".They shut down the machine, unplugged everything and rebooted, but the same error message came up before the start screen would appear therefore forbidding them from entering a case.Another reboot was attempted with the same error message.The procedure was completed with manual instrumentation without significant delays.The patient was not harmed beyond the reported problem.Additionally, they left the navio plugged in for 15 minutes and tried starting it up but the same message came up, forbidding them from entering the system.
|