According to the complaint, on (b)(6) 2022, measurement results for potassium (k+) appeared to be slightly lower in comparison to the results measured by other abls at the same hospital.The discrepant measurement was 2.8 mmol/l and the comparison measurement was 3.4 mmol/l.No reports of death or serious injury.
|
Analysis supports cl parameter ok and that increasing errors on na, k, ca and ph blood measurements as a function of in-use lifetime is seen.In addition, analysis supports the complaint information that the problems are restricted to the operating room / anesthetic recovery room setting.Summing up analysis for this complaint strongly suggest contamination of sensors as root cause.Only cleaning and maintenance procedures specified in instruction for use (ifu) are to be performed.It is suggested to customer to avoid use of any type of cleaning wipes for cleaning analyzer inlet etc.As they may contain surface active substances which damage the sensors.
|