| Model Number |
WW1000EN |
| Medical Device Problem Code |
Unable to Obtain Readings (1516)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
One unit was returned for investigation.Upon physical inspection, it was found that the complained issue could be duplicated.A finger probe with internal wire break resulting in open connection caused the reported problem.The problem source was unable to be identified.A dhr review is not applicable in this case because the defective spo2 probe is not evaluated or tested in any way during the manufacturing process.
|
| |
|
Event or Problem Description
|
|
It was reported that the customer has trouble getting readings.It is not holding a signal well.When we do get a reading it does not seem to be accurate.
|
| |
|