Model Number PS10.0-65D250
Medical Device Problem Codes Corroded (1131); Detachment of Device or Device Component (2907); Naturally Worn (2988)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/18/2022
Type of Reportable Event Malfunction
Event or Problem Description
Information was received that upon explant of the nail after it had achieved intended therapeutic lengthening, it was noted the nail exhibited crevice corrosion and wear between the nail and fixation screws.Additionally, it was found that the crown of the nail had become displaced and appeared to have cold-welded to the male component instead of staying within the female component.No patient harm or adverse impact was reported.
 
Additional Manufacturer Narrative
The device has not been returned for evaluation.The root cause is unable to be determined at this time.If any additional information is provided, a supplemental report will be submitted.
 
Additional Manufacturer Narrative
Device evaluation: during evaluation of the returned nail it was observed that the anti-rotation lug (crown) was disengaged from the housing tube and stuck to the distraction rod.Visual inspection also revealed that the distraction rod of the nail had crevice corrosion.The work order was reviewed and confirmed the device passed all inspections per the acceptance test and the manufacturing x-ray image showed the anti-rotation lug was seated properly in the housing body.The cause for the disengagement of the anti-rotation lug from the housing body is crevice corrosion and subsequent wear of the mating features of the anti-rotation lug and the housing tube, which loosened the assembly between the lug and the housing tube.When load was applied to the nail during patient activities, the loose assembly caused the lug to disassociate from the housing tube.A corrective action is in progress to further investigate the root cause of the crevice corrosion.Device history records review: review of the device history records (dhr) indicated the unit met all quality specifications and testing prior to release.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRECICE STRYDE SYSTEM
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
NUVASIVE SPECIALIZED ORTHOPEDICS, INC.
101 enterprise dr, ste 100
aliso viejo CA 92656
Manufacturer (Section G)
NUVASIVE SPECIALIZED ORTHOPEDICS, INC.
101 enterprise dr, ste 100
aliso viejo CA 92656
Manufacturer Contact
geoff gannon
101 enterprise dr, ste 100
aliso viejo, CA 92656
8583448112
MDR Report Key13367897
Report Number3006179046-2022-00117
Device Sequence Number11760385
Product Code HSB
UDI-Device Identifier00887517951175
UDI-Public887517951175
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K180503
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberPS10.0-65D250
Device Lot Number9041810AAA
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/18/2022
Supplement Date Received by Manufacturer03/30/2022
Initial Report FDA Received Date01/27/2022
Supplement Report FDA Received Date04/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/18/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexFemale
-
-