• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NUVASIVE SPECIALIZED ORTHOPEDICS, INC. PRECICE STRYDE SYSTEM; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NUVASIVE SPECIALIZED ORTHOPEDICS, INC. PRECICE STRYDE SYSTEM; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Model Number PS10.0-65D250
Medical Device Problem Codes Corroded (1131); Detachment of Device or Device Component (2907); Naturally Worn (2988)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/18/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device has not been returned for evaluation.The root cause is unable to be determined at this time.If any additional information is provided, a supplemental report will be submitted.
 
Event or Problem Description
Information was received that upon explant of the nail after it had achieved intended therapeutic lengthening, it was noted the nail exhibited crevice corrosion and wear between the nail and fixation screws.Additionally, it was found that the crown of the nail had become displaced and appeared to have cold-welded to the male component instead of staying within the female component.No patient harm or adverse impact was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRECICE STRYDE SYSTEM
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
NUVASIVE SPECIALIZED ORTHOPEDICS, INC.
101 enterprise dr, ste 100
aliso viejo CA 92656
Manufacturer (Section G)
NUVASIVE SPECIALIZED ORTHOPEDICS, INC.
101 enterprise dr, ste 100
aliso viejo CA 92656
Manufacturer Contact
geoff gannon
101 enterprise dr, ste 100
aliso viejo, CA 92656
8583448112
MDR Report Key13367931
Report Number3006179046-2022-00116
Device Sequence Number19842085
Product Code HSB
UDI-Device Identifier00887517951175
UDI-Public887517951175
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K180503
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 01/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberPS10.0-65D250
Device Lot Number9041810AAA
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/18/2022
Initial Report FDA Received Date01/27/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/18/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexFemale
-
-