| Catalog Number |
110003452 |
| Medical Device Problem Code |
Difficult or Delayed Separation (4044)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/04/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.Concomitant medical products: item # 010002736/ g7 ball hex drvr for insr hndl/ lot # 062100; item # 110010247/ g7 osseoti 4 hole shell 58mm/ lot # unknown; item # 110003451/ g7 str modular shell inserter/ lot # 494810.Multiple mdr reports were filed for this event, please see associated reports: 0001825034 -2022 -00195, 0001825034 -2022 -00197 and 0001825034 -2022 -00199.
|
| |
|
Event or Problem Description
|
|
It was reported that during a hip procedure the cup insertion handle could not be loosened from cup.Tommy bar was used with g7 ball tip screw driver to try to release cup.G7 ball driver broke at tip and insertion handle was still attached to cup in patient.Whole construct was removed.No adverse events were reported.Attempts have been made and additional information on the reported event is unavailable at this time.
|
| |
|
Event or Problem Description
|
|
No additional event information to report at this time.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.Visual inspection of the returned threaded shaft found wear rings and scratching on the shaft.The large hex feature is dinged and scratched.The lead thread on the tip has been deformed.Dhr was reviewed and no discrepancies were found.The root cause is unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|