• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION XXL VASCULAR; DILATOR, ESOPHAGEAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION XXL VASCULAR; DILATOR, ESOPHAGEAL Back to Search Results
Model Number 23375
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/16/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that balloon rupture occurred.The target lesion was located in the iliac vein.A 14-4/5.8/75 xxl balloon dilator was advanced for dilatation.However, during inflation before reaching the nominal pressure, the balloon ruptured.The procedure was completed with another of the same device.No patient complications were reported and the patient status was stable.
 
Event or Problem Description
It was reported that balloon rupture occurred.The target lesion was located in the iliac vein.A 14-4/5.8/75 xxl balloon dilator was advanced for dilatation.However, during inflation before reaching the nominal pressure, the balloon ruptured.The procedure was completed with another of the same device.No patient complications were reported and the patient status was stable.It was further reported that the target lesion was 30% stenosed, moderately tortuous and severely calcified.The balloon ruptured during initial inflation.
 
Additional Manufacturer Narrative
Device evaluated by mfr: the xxl/14-4/5.8/75 was received for analysis.A visual examination identified that the balloon was not folded which indicates that the balloon was subjected to positive pressure.Blood present in balloon which is indicative of a leak.The balloon was attached to an encore inflation device, subjected to positive pressure and liquid was observed to be leaking from a balloon pinhole located 13mm distal from the distal markerband.An examination of the balloon material and the tip region identified no issues which could potentially have contributed to this complaint.A visual and tactile examination identified no kinks or damage to the shaft which may have potentially contributed to the complaint incident.A visual examination identified no issues with the tip of the device that could have contributed to the complaint incident.
 
Event or Problem Description
It was reported that balloon rupture occurred.The target lesion was located in the iliac vein.A 14-4/5.8/75 xxl balloon dilator was advanced for dilatation.However, during inflation before reaching the nominal pressure, the balloon ruptured.The procedure was completed with another of the same device.No patient complications were reported and the patient status was stable.It was further reported that the target lesion was 30% stenosed, moderately tortuous and severely calcified.The balloon ruptured during initial inflation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XXL VASCULAR
Common Device Name
DILATOR, ESOPHAGEAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key13368709
Report Number2134265-2022-00787
Device Sequence Number19821550
Product Code LIT
UDI-Device Identifier08714729190462
UDI-Public08714729190462
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K150303
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/12/2024
Device Model Number23375
Device Catalogue Number23375
Device Lot Number0027129803
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/09/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 01/20/2022
Supplement Date Received by Manufacturer01/26/2022
03/03/2022
Initial Report FDA Received Date01/27/2022
Supplement Report FDA Received Date02/07/2022
03/09/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/13/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-