• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. TM REVERSE STEM 12MM X 130MM; PROSTHESIS, SHOULDER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. TM REVERSE STEM 12MM X 130MM; PROSTHESIS, SHOULDER Back to Search Results
Model Number 00-4349-012-13
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 01/17/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient underwent a revision surgery approximately four years post initial due to disassociation of poly liner from the stem.The loose liner was explanted and a new poly liner was implanted.No other implants were revised.No additional patient consequences were reported.
 
Additional Manufacturer Narrative
(b)(4).Concomitant medical products: cat: 00434904006, lot: 62977613, 40mm poly liner plus 6mm ofs; cat: 00434901500, lot: 63715593, 15mm post length base plate; cat: 00434904011, lot: 63653808, tm reverse 40mm glenosphere; cat: 01.04223.036, lot: 2921546, invers/revers scr syst 4.5- 36; cat: 01.04223.036, lot: 2921529, invers/revers scr syst 4.5- 36.Foreign- (b)(6).Customer has indicated that the product will not be returned to zimmer biomet for investigation, as the device remains implanted.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0001822565-2022-00233.
 
Additional Manufacturer Narrative
Reported event was confirmed by review of radiographs.Review of the available radiographs identified that left total shoulder reversed arthroplasty with possible loosening of the glenoid component with the associated changing orientation of the glenosphere as well as polyethylene wear of the glenohumeral component.Device history record (dhr) was reviewed and no discrepancies were found.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No additional event information is available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TM REVERSE STEM 12MM X 130MM
Common Device Name
PROSTHESIS, SHOULDER
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key13368745
Report Number0001822565-2022-00234
Device Sequence Number3991787
Product Code KWT
UDI-Device Identifier00889024268951
UDI-Public(01)00889024268951(17)270831(10)63700625
Combination Product (Y/N)N
Initial Reporter CountryDA
PMA/510(K) Number
K052906
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number00-4349-012-13
Device Catalogue Number00434901213
Device Lot Number63700625
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/17/2022
Supplement Date Received by Manufacturer03/28/2022
Initial Report FDA Received Date01/27/2022
Supplement Report FDA Received Date04/06/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/18/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age77 YR
Patient SexMale
Patient Weight117 KG
-
-