This follow-up report is submitted to fda in accord with applicable regulations ¿ and as indicated by terumo cardiovascular systems in the initial report submitted to the fda on january 27, 2022.Upon further investigation of the reported event, the following information is new and/or changed: d4 (additional device information - added expiration date) g3 (date received by manufacturer) g6 (indication that this is a follow-up report) h2 (follow-up due to additional information) h4 (device manufacture date) h6 (identification of evaluation codes 3331, 4114, 4246, 4308) type of investigation #1: 3331 - analysis of production records type of investigation #2: 4114 - device not returned investigation finding: 4246 - transport/storage problem identified investigation conclusions: 4308 - cause traced to transport/storage the investigation verified that incorrect product was shipped.An sncr was sent for this incident to the appropriate distribution center for further investigation.Based on the sncr investigation, the root cause for this incident was the distribution center picked and sent incorrect product.All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
|