| Model Number |
300800 |
| Medical Device Problem Code |
No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/28/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Results which were positive for the neumodx sars-cov-2 test strip on the on the neumodx 288 molecular system were discrepant with another method.
|
| |
|
Additional Manufacturer Narrative
|
|
No adverse outcomes were reported.The customer routinely confirms results using other assay platforms.We are reporting this event in an abundance of caution and in accordance with the eua requirements.
|
| |
|
Search Alerts/Recalls
|