• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 POWERPICC PROVENA SOLO 5F TL MAX BARRIER PLUS KIT; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 POWERPICC PROVENA SOLO 5F TL MAX BARRIER PLUS KIT; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS Back to Search Results
Model Number N/A
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/11/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The manufacturer has received the sample and will evaluate.Results are expected soon.A lot history review (lhr) of refw2578 showed no other similar product complaint(s) from this lot number.
 
Event or Problem Description
It was reported "completed dressing change and when i flushed it no blood return and difficult to flush.Cathflow no results.When flushing felt like it was malpositioned.Obtain x-ray.Catheter appeared to be kinked in the subclavian.".
 
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, complaint and batch history, applicable previous investigation(s), applicable manufacturing records, radiographic image analysis and applicable fmea documents.Based on a review of this information, the following was concluded: the complaint of a kink in the catheter was inconclusive due to the sample condition.A chest radiograph, that was a coned down frontal projection, was returned for evaluation.A right-sided catheter was present in the image.The image was forwarded to a radiologist consultant for further review.As per their review, the image quality was poor and difficult to interpret.There were overlapping areas in the catheter, but a kink could not be confirmed in the image.Tactile evaluation, functional testing, and dimensional analysis could not be conducted without the physical sample.Although a kink could not be conclusively confirmed, possible contributing factors for a kinked catheter include the placement technique and/or patient anatomy.The ifu states, ¿avoid placement or securement of the catheter where kinking may occur, to minimize stress on the catheter, patency problems or patient discomfort.¿ and ¿avoid sharp or acute angles during implantation which could compromise the patency of the catheter lumen.¿ the event description stated that the catheter could not be flushed or aspirated.Additional contributing factors for an occluded catheter that is not kinked could include buildup of biological material and/or precipitate formation.H3 other text : evaluation findings are in section h.11.
 
Event or Problem Description
It was reported "completed dressing change and when i flushed it no blood return and difficult to flush.Cathflow no results.When flushing felt like it was malpositioned.Obtain x-ray.Catheter appeared to be kinked in the subclavian.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERPICC PROVENA SOLO 5F TL MAX BARRIER PLUS KIT
Common Device Name
CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
kelsey erickson
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key13373636
Report Number3006260740-2022-00099
Device Sequence Number17086046
Product Code LJS
UDI-Device Identifier00801741155024
UDI-Public(01)00801741155024
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K072230
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue NumberS1395108D3
Device Lot NumberREFW2578
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/20/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/11/2022
Supplement Date Received by Manufacturer03/07/2022
Initial Report FDA Received Date01/27/2022
Supplement Report FDA Received Date03/11/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-