• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 EZ HUBER 20G X 15MM; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 EZ HUBER 20G X 15MM; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Model Number N/A
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/10/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The date of event was not provided by the complainant/reporter, the date reflected in this report is the date bd became aware of the event.The device has not been returned to the manufacturer for evaluation.A lot history review (lhr) of reez3291 showed two other similar product complaint(s) from this lot number.
 
Event or Problem Description
It was reported that on withdrawing blood, leak discovered where port tubing meets hub.No harm to patient or staff.It was stated the needle removed and replaced with a new one.
 
Additional Manufacturer Narrative
H11: section a through f the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation:¿patient severity, applicable previous investigation(s), sample analysis, applicable fmea documents, applicable manufacturing records, and labeling.Based on a review of this information, the following was concluded: the complaint of a leak at the needle housing is confirmed but the exact cause remains unknown.One 20 ga ez huber infusion set was returned for evaluation.The product label returned indicated lot: reez3291.The safety mechanism was returned fully engaged over the needle tip.The infusion set was flushed with water and fluid was observed to leak at the clear plastic needle housing.Microscopic observation of the leak location revealed multiple horizontal cracks in the clear housing material.The cracks appeared to have propagated from the proximal end of the housing up to the region where the needle is located.The cracks also appeared to be aligned in the same region as the location of the extension tubing within the housing.Based on the condition of the returned sample, possible contributing factors include damage during storage and handling.Since the location at which the component may have been exposed to external stresses could not be determined, the complaint is confirmed but the exact cause remains unknown at this time.H3 other text : evaluation findings are in section h.11.
 
Event or Problem Description
It was reported that on withdrawing blood, leak discovered where port tubing meets hub.No harm to patient or staff.It was stated the needle removed and replaced with a new one.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EZ HUBER 20G X 15MM
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
kayla olsen
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key13374009
Report Number3006260740-2022-00101
Device Sequence Number5011110
Product Code FPA
UDI-Device Identifier00801741089671
UDI-Public(01)00801741089671
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K040533
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue NumberSHN20-60
Device Lot NumberREEZ3291
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/10/2022
Supplement Date Received by Manufacturer09/26/2022
Initial Report FDA Received Date01/27/2022
Supplement Report FDA Received Date09/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/01/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-