AMO MANUFACTURING NETHERLANDS TECNIS IOL; INTRAOCULAR LENS Back to Search Results
Model Number DFR00V
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Corneal Edema (1791); Flashers (1864); Vitreous Floaters (1866); Visual Impairment (2138); Visual Disturbances (2140); Halo (2227)
Date of Event 07/19/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The intraocular lens (iol) was not returned for evaluation; therefore, a failure analysis of the complaint device cannot be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.All pertinent information available to johnson & johnson surgical vision, inc has been submitted.
 
Event or Problem Description
It was reported that an intraocular lens (iol) was explanted in the right eye (od) due to distance vision was not good and near vision is ok, the worst at night, with bad glare and halos.New cobweb floaters reported 2 weeks after surgery and flashes were very frequent and very distracting.Vision is still foggy.Patient feels very uncomfortable driving.Pre-operative (op) was 20/40 and post-op is 20/40.Mild corneal swelling remains, no edema on mac optical coherence tomography (oct).It was discussed with the patient that blurred visual acuity (va) is anticipated to improve as eye heals with residual nearsightedness noted.There was an incision enlargement during the explant.There was no patient injury.Another johnson & johnson dib00 lens (different model and lower diopter) was implanted as a replacement.The patient is doing great post-operatively.No additional information was provided.
 
Additional Manufacturer Narrative
Additional information: section d9.Device available for evaluation? yes.Returned to manufacturer on feb 1, 2022.Section h3.Device evaluated by manufacturer? yes.Device evaluation: the product was returned to the manufacturing site for evaluation.Visual inspection under magnification revealed that the complaint lens was received cut in half, which is consistent with a lens that was handled during explant.The lens was cleaned, and no issues related to or could have contributed to visual issues were observed.Based on the return condition of the lens, the complaint issue could not be confirmed, and no product deficiency could be identified.Manufacturing record review: the manufacturing records for the product were reviewed.The product was manufactured and released according to specification.A search of complaints related to this production order (po) was performed.The search revealed one additional complaint folder was received for this po.However, complaint issues reported are not related.Therefore, no escalations are required.Conclusion: as a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
AMO MANUFACTURING NETHERLANDS
van swietenlaan 5
groningen, groningen 9728 NX
NL  9728 NX
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key13375056
Report Number3012236936-2022-00248
Device Sequence Number14717248
Product Code HQL
UDI-Device Identifier05050474750753
UDI-Public(01)05050474750753(17)240411
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/11/2024
Device Model NumberDFR00V
Device Catalogue NumberDFR00VU240
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/05/2022
Supplement Date Received by Manufacturer03/08/2022
Initial Report FDA Received Date01/27/2022
Supplement Report FDA Received Date03/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/11/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-