Model Number 9500 |
Device Problem
Display or Visual Feedback Problem (1184)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/01/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Attempts have been made to obtain the product.The product has not been returned to masimo to allow an analysis to be performed.If the product is returned for evaluation or new information is obtained, a follow up report will be submitted.
|
|
Event Description
|
The customer reported the device screen freezes.No consequences or impact to patient were reported.
|
|
Event Description
|
The customer reported the device screen freezes.No consequences or impact to patient were reported.
|
|
Manufacturer Narrative
|
Additional manufacturing narrative: other, other text: the returned device was evaluated.Visual inspection upon receiving revealed foreign contaminant observed on the touchscreen surface.During operational testing, abnormal touchscreen functionality was observed.The touchscreen responded to physical touch, but the screen was also activated without physical touch.Temporarily installed a touchscreen from known good radical-7 and customer device was fully functional.A service history record review reveals that this unit was in the field for over seven (7) years with no previous reported issues related to this reported event.
|
|