• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AMO PUERTO RICO MFG. INC. TECNIS IOL; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMO PUERTO RICO MFG. INC. TECNIS IOL; INTRAOCULAR LENS Back to Search Results
Model Number ZA9003
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/09/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the intraocular lens (iol) had bent/ broken haptic.There was contact with patient's right eye and the issue was noticed while inserting the lens into the eye.The procedure was completed using another lens of same model and diopter and the patient has recovered.No further information was available.
 
Additional Manufacturer Narrative
Age: unknown, information not provided.Patient weight: unknown, information not provided.Ethnicity/race: unknown, information not provided.If implanted, give date: unknown, information not provided.If explanted, give date: unknown, information not provided.Device evaluation: product testing could not be performed since the product was not returned for evaluation.Therefore, the reported issue could not be verified, and product quality deficiency could not be determined.Manufacturing records review: the manufacturing records for the intraocular lens were reviewed and no non-conformance report (nc) was found as part of this manufacturing records review.The product was manufactured and released according to specification.A search in complaint system revealed that one additional complaint was received from this production order which was voided and no escalation was required.Conclusion: as a result of the investigation, there is no indication of a product quality deficiency.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
Section d9: device available for evaluation: yes.Section d9: returned to manufacturer on: feb 2, 2022.Device evaluation: the product was returned to the manufacturing site for evaluation.Visual inspection under magnification revealed viscoelastic residue on the optic body and haptics.The lens was cleaned, and haptic damage was observed.No further issues were identified.The complaint issue could be confirmed; however, this may be attributed to handling during insertion, and cannot be confirmed to be related to a manufacturing issue.Therefore, no product deficiency could be identified.Conclusion: as a result of the investigation there is no indication of a product quality deficiency.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key13375909
Report Number3012236936-2022-00235
Device Sequence Number20069975
Product Code HQL
UDI-Device Identifier05050474529076
UDI-Public(01)05050474529076(17)250531
Combination Product (Y/N)N
Initial Reporter StateNE
Initial Reporter CountryUS
PMA/510(K) Number
P990080
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2025
Device Model NumberZA9003
Device Catalogue NumberZA90030230
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/03/2022
Supplement Date Received by Manufacturer03/11/2022
Initial Report FDA Received Date01/27/2022
Supplement Report FDA Received Date03/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/31/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
EMERALDC30 CARTRIDGE, LOT UNKNOWN
Patient SexFemale
-
-