• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION INCEPTA CRT-D; CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION INCEPTA CRT-D; CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR Back to Search Results
Model Number N164
Medical Device Problem Codes High impedance (1291); Pacing Problem (1439); Inappropriate/Inadequate Shock/Stimulation (1574); Under-Sensing (1661); Noise, Audible (3273)
Health Effect - Clinical Code Electric Shock (2554)
Date of Event 12/05/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) inappropriately delivered anti-tachycardia pacing (atp) and six shocks due to supraventricular tachycardia (svt).The ventricular rhythm surpassed the atrial rhythm due to cross chamber blanking.This crt-d remains in service.No additional adverse patient effects were reported.
 
Event or Problem Description
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) inappropriately delivered anti-tachycardia pacing (atp) and six shocks due to supraventricular tachycardia (svt).The ventricular rhythm surpassed the atrial rhythm due to cross chamber blanking.This crt-d remains in service.No additional adverse patient effects were reported.Additional information was received and this crt-d is exhibiting beeping.Technical services (ts) was consulted and found high out of range shock lead impedance measurements.Technical services (ts) recommended to continue to monitor the device.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INCEPTA CRT-D
Common Device Name
CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key13376321
Report Number2124215-2022-01942
Device Sequence Number3452839
Product Code LWP
UDI-Device Identifier00802526481031
UDI-Public00802526481031
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
P010012/S255
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date01/21/2016
Device Model NumberN164
Device Catalogue NumberN164
Device Lot Number105486
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/06/2021
Supplement Date Received by Manufacturer04/21/2022
Initial Report FDA Received Date01/27/2022
Supplement Report FDA Received Date05/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/03/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient Age58 YR
Patient SexFemale
-
-