| Model Number |
Z9002 |
| Medical Device Problem Codes |
Inaccurate Delivery (2339); Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/11/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Age: unknown, information not provided.Ethnicity/race: unknown, information not provided.If implanted, give date: not applicable, as there is no indication that the lens was implanted if explanted, give date: not applicable, as there is no indication that the lens was implanted; therefore, not explanted.The device was not returned for analysis.Therefore, a failure analysis of the complaint device cannot be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the intraocular lens (iol) was partially inserted into the patient's left eye, however, it was removed as the lens was torn.Another lens of same model and diopter was used as the replacement.There was no injury and no other intervention was required.The patient was fine post operation.No further information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Section d9: device available for evaluation: yes.Section d9: returned to manufacturer on: feb 4, 2022.Section h3: device evaluated by manufacturer: yes.Device evaluation: the product was returned to the manufacturing site for evaluation.Visual inspection under magnification revealed that the lens was received with a torn optic body and haptic damage.The complaint issue, delivery issue, could not be confirmed.The complaint issue, iol torn, could be confirmed; however, it may be attributed to handling during implantation of the lens, and cannot be confirmed to be related to a manufacturing issue.Therefore, no product deficiency could be identified.Manufacturing record review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.The search revealed one additional investigation request (ir) for this production order number has been received, however, was not related to this complaint issue.Conclusion: as a result of the investigation there is no indication of a product quality deficiency.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Search Alerts/Recalls
|
|