• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN MFG DC BOULDER SONICISION; INSTRUMENT, ULTRASONIC SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN MFG DC BOULDER SONICISION; INSTRUMENT, ULTRASONIC SURGICAL Back to Search Results
Model Number SCD391
Medical Device Problem Codes Difficult or Delayed Activation (2577); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code Device Embedded In Tissue or Plaque (3165)
Date of Event 12/27/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during a procedure, the device was difficult to activate.A new dissector was used with the same generator and battery and it worked fine.
 
Additional Manufacturer Narrative
Additional information: b1, b2(removed), b5, g3, h1, h3, h6 new information has been received pertaining to the event.This event has been reassessed and the reportability has been determined to be a malfunction.H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the waveguide to be broken.Functional testing found the assembled device returned a green light and produced a welcome tone and immediately alarmed when activated.The waveguide was inspected under magnification and the origin of the crack was identified.It was also concluded that the titanium waveguide became cracked from continued use after it may have come in contact with a metallic object.The crack propagated until the waveguide fractured.No defects or damage were observed in the packaging returned that could have contributed to the reported defect.It was reported that the device was difficult to activate.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.The instructions included with this device provide the following guidance: contact between the active blade and other metal objects (hemostats, clips, staples, retractors, etc.) may result in unintended damage to tissue and/or device failure.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during a procedure, the device was difficult to activate.A new dissector was used with the same generator and battery and it worked fine.Medtronic initial evaluation found that the disposable device revealed that the tip of the waveguide had fractured and disengaged.The broken piece was not returned.The piece did not fall into patient's cavity.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SONICISION
Common Device Name
INSTRUMENT, ULTRASONIC SURGICAL
Manufacturer (Section D)
COVIDIEN MFG DC BOULDER
5920 longbow dr
boulder CO 80301 3299
Manufacturer (Section G)
COVIDIEN MFG DC BOULDER
5920 longbow dr
boulder CO 80301 3299
Manufacturer Contact
tracy landers
5920 longbow drive
boulder, CO 80301
3035816943
MDR Report Key13382771
Report Number1717344-2022-00104
Device Sequence Number5184662
Product Code LFL
UDI-Device Identifier10884521052659
UDI-Public10884521052659
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K101797
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/13/2026
Device Model NumberSCD391
Device Catalogue NumberSCD391
Device Lot Number10050205X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/11/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/11/2022
Supplement Date Received by Manufacturer02/11/2022
Initial Report FDA Received Date01/28/2022
Supplement Report FDA Received Date02/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/14/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-