Model Number 2090
Medical Device Problem Codes Crack (1135); No Display/Image (1183)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/10/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the programmer display screen was responsive to the stylus only in the left lower corner of the screen and the display screen was cracked with shards of glass observed.There was no patient involvement.
 
Additional Manufacturer Narrative
Product analysis: analysis was able to confirm the customer comment that the programmer display screen was only responsive to the stylus in the left lower corner of the screen, the display screen was cracked with shards of glass observed.The overlay was replaced.The nylon standoff between link electronic module (lem) and the micro processor unit (mpu) were replaced as a preventative measure.The hard drive was reconfigured, reloaded and the software was updated.The device passed all final functional tests.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARELINK PROGRAMMER
Common Device Name
PROGRAMMER, PACEMAKER
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer (Section G)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key13384640
Report Number2182208-2022-00284
Device Sequence Number3452496
Product Code KRG
UDI-Device Identifier00613994434937
UDI-Public00613994434937
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
P890003
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 02/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number2090
Device Catalogue Number2090
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/12/2022
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/10/2022
Supplement Date Received by Manufacturer02/04/2022
Initial Report FDA Received Date01/28/2022
Supplement Report FDA Received Date02/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/29/2009
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
2067L RADIOFREQUENCY HEAD, 2290 ANALYZER
Patient SexUnknown
-
-