Model Number 84592
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/06/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(6).
 
Event or Problem Description
It was reported that the device was fractured upon removal.The target lesion was located in the severely tortuous and moderately calcified artery in the knee area.A direxion fathom-16 system was selected for use.During the procedure, both non boston scientific guidewire and the catheter became very jumbled up.The device was hydrated prior to removal of the device.When the device was removed completely, it was noted that the catheter was broken 3 inches from the tip of the device.No patient complications were reported.
 
Event or Problem Description
It was reported that the device was fractured upon removal.The target lesion was located in the severely tortuous and moderately calcified artery in the knee area.A direxion fathom-16 system was selected for use.During the procedure, both non boston scientific guidewire and the catheter became very jumbled up.The device was hydrated prior to removal of the device.When the device was removed completely, it was noted that the catheter was broken 3 inches from the tip of the device.No patient complications were reported.
 
Additional Manufacturer Narrative
E1- initial reporter address 1: (b)(6).Device evaluated by manufacturer: the device was returned for analysis.The device shaft was analyzed for any damage.The devices nickel shaft showed damage in the form of a fracture located 126cm from the hub.The fracture looked to be consistent with a tensile force breaking as the device was not completely separated, the inner liner was still attached.Inspection of the remainder of the device, apart from the observed damage, revealed no other damage or irregularities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DIREXION FATHOM-16 SYSTEM
Common Device Name
CATHETER, CONTINUOUS FLUSH
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORK LIMITED
model farm road
cork
EI  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key13384683
Report Number2134265-2022-00348
Device Sequence Number14372481
Product Code KRA
UDI-Device Identifier08714729839637
UDI-Public08714729839637
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K142259
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/27/2023
Device Model Number84592
Device Catalogue Number84592
Device Lot Number0028275601
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/26/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 01/06/2022
Supplement Date Received by Manufacturer01/28/2022
Initial Report FDA Received Date01/28/2022
Supplement Report FDA Received Date02/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/27/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexMale
-
-