• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL BRK-1¿ TRANSSEPTAL NEEDLE, 71 CM LENGTH; TROCAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL BRK-1¿ TRANSSEPTAL NEEDLE, 71 CM LENGTH; TROCAR Back to Search Results
Model Number 407201
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Cardiac Perforation (2513)
Date of Event 01/25/2022
Type of Reportable Event Serious Injury
Event or Problem Description
During an atrial fibrillation procedure, a perforation occurred during the transseptal puncture near the fossa ovalis.Moreover, when attempting the transseptal puncture, a high amount of resistance was felt and the tissue became colored with contrast.The patient then became hypotensive and a pericardial effusion was diagnosed via transthoracic echocardiogram.The procedure was then cancelled and heparin was administered to the patient.The patient remained stable and no pericardiocentesis was performed.The patient was then sent to the intensive care unit for monitoring.There were no performance issues with the abbott device.
 
Additional Manufacturer Narrative
Correction: h6.
 
Additional Manufacturer Narrative
Correction: h6 (health effect - impact code).
 
Additional Manufacturer Narrative
Additional information: d9, g3, h2, h3, h6.One brk transseptal needle/stylet assembly were received for evaluation.Functional testing revealed no resistance or functional anomalies were noted when the returned brk needle/stylet assembly was advanced through a dilator/sheath assembly from current inventory.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the reported cardiac perforation remains unknown.Per the ifu, cardiac perforation is a known risk during the use of this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BRK-1¿ TRANSSEPTAL NEEDLE, 71 CM LENGTH
Common Device Name
TROCAR
Manufacturer (Section D)
ST. JUDE MEDICAL
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS  1897-4050
Manufacturer (Section G)
ST. JUDE MEDICAL
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS   1897-4050
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key13385214
Report Number3008452825-2022-00030
Device Sequence Number13914142
Product Code DRC
UDI-Device Identifier05414734205160
UDI-Public05414734205160
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K072278
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 03/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/31/2024
Device Model Number407201
Device Catalogue Number407201
Device Lot Number8150606
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/16/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/28/2022
Supplement Date Received by Manufacturer01/28/2022
02/04/2022
03/08/2022
Initial Report FDA Received Date01/28/2022
Supplement Report FDA Received Date01/28/2022
02/04/2022
03/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/24/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age68 YR
Patient SexFemale
Patient Weight55 KG
-
-