As reported, during the procedure of a 23 mm sapien 3 ultra valve in the aortic position via transfemoral approach, there was difficulty inserting the esheath into the patient.A peripheral angioplasty of the calcium was performed and the angioplasty balloon was left inflated inside the tip of the esheath.This action damaged the tip of the esheath causing pre-expansion the tip which was noted to be split and frayed.There was no injury to the patient.
|
|
A supplemental mdr is being submitted for correction and additional information base on the device evaluation.The following sections of this report have been updated: corrected h.6 component codes and investigation conclusions and added new information to h.6 type of investigation and investigation findings.The esheath was not returned to edwards lifesciences for evaluation.Without the device returned for evaluation, visual inspection, functional testing and dimensional analysis were unable to be completed.Imagery was provided from the site and revealed the following: calcification and undersized vessels present in patient's access vasculature.During manufacturing, the device undergoes multiple 100% inspections.In addition, the work order underwent functional product verification testing on a sampling basis as a requirement for lot release.These inspections and tests during the manufacturing process support that it is unlikely that a non-conformance contributed to the reported complaint.The instructions for use/training manuals were reviewed for guidance/instruction involving the esheath and delivery system usage.Based on the review of the ifu/training manuals, no deficiencies were identified.The complaints were unable to be confirmed.As the device was not returned, engineering was unable to perform any visual inspection, functional testing or dimensional analysis.Therefore, the presence of a manufacturing non-conformance was unable to be determined.However, reviews of the dhr, lot history, and complaint history did not provide any indication that a manufacturing non-conformance would have contributed to the complaint event.A review of the ifu and training manuals revealed no deficiencies.Furthermore, no abnormalities were observed during device unpacking or preparation, indicating that the device was not found damage out of the box.In this case, ''there was difficulty inserting the esheath into the patient.A peripheral angioplasty of the calcium was performed and may have left the angioplasty balloon inflated inside the tip of the esheath accidentally.This action may have damaged the tip of the esheath causing pre-expansion the tip which was noted to be split and frayed''.Per the training manual, ''push force can vary due to angle of insertion, vessel diameter, tortuosity and degree of calcification''.The presence of calcification can create a challenging pathway for sheath insertion as it can create friction through contact with sheath shaft.Additionally, as the angioplasty balloon may have been inflated within the sheath, the expansion of the balloon likely pushed outward against the sheath distal tip and split it.As such, available information suggests that patient (calcification, tortuosity) and procedural factors (balloon inflated within sheath) may have contributed to the complaint events.Complaint histories for all reported events are reviewed against trending control limits monthly, and any excursions above the control limits are assessed and documented as part of this monthly review.No ifu, labeling, and training manual inadequacies were identified.Therefore, no corrective or preventative action (capa) nor product risk assessment (pra) is required at this time.
|