• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC TORIC INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC TORIC INTRAOCULAR LENS Back to Search Results
Model Number TMICL13.2
Medical Device Problem Code Inadequacy of Device Shape and/or Size (1583)
Health Effect - Clinical Codes Intraocular Pressure Increased (1937); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 01/03/2022
Type of Reportable Event Serious Injury
Event or Problem Description
The reporter indicated that the surgeon implanted a 13.2mm tmicl13.2 lens, -15.0/+1.0/036 (sphere/cylinder/axis) into the patient's left (os) eye on (b)(6) 2022.On (b)(6) 2022 the lens was exchanged with a shorter lens due to excessive vault, elevated iop, and narrow angles.The problem is resolved.
 
Additional Manufacturer Narrative
Weight: unk.Ethnicity: unk.Race: unk.Claim#: (b)(4).
 
Additional Manufacturer Narrative
Weight: unk.Ethnicity: unk.Race: unk.(b)(4).Claim#: (b)(4).
 
Event or Problem Description
The reporter indicated that the surgeon implanted a 13.2mm tmicl13.2 lens, -15.0/+1.0/036 (sphere/cylinder/axis) into the patient's left (os) eye on (b)(6) 2022.On (b)(6) 2022 the lens was exchanged with a shorter lens due to excessive vault, elevated iop, and narrow angles.The problem is resolved.
 
Additional Manufacturer Narrative
Reportedly the lens was too long, causing a shallow angle and iop issues.Claim# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Common Device Name
PHAKIC TORIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer (Section G)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer Contact
cynthia thai
1911 walker avenue
monrovia, CA 91016
8002927902
MDR Report Key13385990
Report Number2023826-2022-00337
Device Sequence Number4858512
Product Code QCB
UDI-Device Identifier00841542109710
UDI-Public00841542109710
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
P030016
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/31/2023
Device Model NumberTMICL13.2
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/05/2022
Supplement Date Received by Manufacturer02/08/2022
Initial Report FDA Received Date01/28/2022
Supplement Report FDA Received Date03/11/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/10/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNK.
Outcome Attributed to Adverse Event Required Intervention;
Patient Age35 YR
Patient SexFemale
-
-