Model Number TGMR373710
Medical Device Problem Code Separation Failure (2547)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/03/2022
Type of Reportable Event Malfunction
Event or Problem Description
On (b)(6) 2022, the patient was treated for a thoracic aortic aneurysm.The physician used an elephant trunk technique using a gore® tag® conformable thoracic stent graft with active control system.The device was deployed, but was unable to be removed from the catheter - (1600-e, catheter events - other/unknown).The lock would not release, and the physician was unaware of the ability to use the escape hatch to resolve the issue, so he decided to use a different gore® tag® conformable thoracic stent graft with active control system.The patient tolerated the procedure.The device will be returned for evaluation.
 
Additional Manufacturer Narrative
H.6.Investigation findings: code 3233 updated to code 213.H.6.Investigation conclusions: code 11 updated to code 4315.The gore® tag® conformable thoracic stent graft with active control system (cmds) device evaluation showed the following: the stent graft was returned attached to the catheter and fully deployed.The lockwire was separated from the lockwire connector and lockwire handle.The lockwire was properly routed through the catheter, device, and into the curved olive.The angulation fibers were separated from the angulation handle but remained routed through the device and catheter.During testing for design verification, process qualification, and ongoing quality control testing, devices are tested in anatomical models representative of cmds indicated use.Based on the evaluation, the reported difficulty and inability to remove the lockwire could not be confirmed.There is potential that off-label use in an antegrade hybrid fashion may have contributed to the reported event.
 
Additional Manufacturer Narrative
Codes for h6 were blank on previous report, although they were reported in h10.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® TAG® CONFORMABLE THORACIC STENT GRAFT WITH ACTIVE CONTROL SYSTEM
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
W. L. GORE & ASSOCIATES, INC.
1505 n. fourth street
flagstaff AZ 86004
Manufacturer (Section G)
MEDICAL WOODY SPRINGS B/P
3450 w. kiltie lane
flagstaff AZ 86005
Manufacturer Contact
greg rawlings
1505 n. fourth street
flagstaff, AZ 86004
9285263030
MDR Report Key13386103
Report Number2017233-2022-02699
Device Sequence Number1817388
Product Code MIH
UDI-Device Identifier00733132646807
UDI-Public00733132646807
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
P040043
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/29/2024
Device Model NumberTGMR373710
Device Catalogue NumberTGMR373710
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 01/03/2022
Supplement Date Received by Manufacturer02/10/2022
02/10/2022
Initial Report FDA Received Date01/28/2022
Supplement Report FDA Received Date02/10/2022
04/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/31/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age64 YR
Patient SexFemale
-
-