• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 ATTUNE CR FB INSRT SZ 6 7MM; ATTUNE IMPLANT : KNEE TIBIAL INSERT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 ATTUNE CR FB INSRT SZ 6 7MM; ATTUNE IMPLANT : KNEE TIBIAL INSERT Back to Search Results
Model Number 1516-20-607
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Muscular Rigidity (1968); Pain (1994)
Date of Event 12/30/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # pc-(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Clinical notification received for revision to address pain and stiffness.Date of implantation: (b)(6) 2015.Date of revision: (b)(6) 2021.(right knee) treatment: revision of tibial tray and tibial insert.
 
Additional Manufacturer Narrative
Product complaint (b)(4).Additional information provided from medical record review, indicates the patient received a revision, due to tibial tray loosening at the cement, to implant interface and migration.It is reasonable to conclude, that the device reported on this medwatch would not cause or contribute to the reported serious injury/revision.As the tibial insert doesn¿t have any interaction between the cement and tibial component.And the tibial insert must be removed to address the loosening.It is also reasonable to conclude, the reported pain and stiffness are secondary to the confirmed revision.The tibial tray component was reported on report number 1818910-2022-01880.At this time depuy synthes joint reconstruction considers, the device on this report to be not reportable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE CR FB INSRT SZ 6 7MM
Common Device Name
ATTUNE IMPLANT : KNEE TIBIAL INSERT
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY RAYNHAM, A DIV. OF DEPUY ORTHO 1219655
325 paramount drive
raynham MA 02767
Manufacturer Contact
kara ditty-bovard
700 orthopaedic dr.
warsaw, IN 46581-0988
6107428552
MDR Report Key13387275
Report Number1818910-2022-01879
Device Sequence Number20058056
Product Code JWH
UDI-Device Identifier10603295045786
UDI-Public10603295045786
Combination Product (Y/N)N
Initial Reporter StateKY
Initial Reporter CountryUS
PMA/510(K) Number
K101433
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Study,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number1516-20-607
Device Catalogue Number151620607
Device Lot Number240604
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/19/2022
Supplement Date Received by Manufacturer02/25/2022
Initial Report FDA Received Date01/28/2022
Supplement Report FDA Received Date03/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/21/2012
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ATTUNE C/RET FEM RT SZ 6 CEMENTED; ATTUNE FB TIB BASE SZ 4 CEM; ATTUNE MEDIAL DOME PAT 35MM
Outcome Attributed to Adverse Event Required Intervention;
Patient Age71 YR
Patient SexFemale
Patient Weight126 KG
-
-