Model Number HST III SYSTEM (4.3MM)
Device Problem Crack (1135)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/28/2021
Event Type  malfunction  
Event Description
The hospital reported that during a coronary artery bypass procedure, hst iii system (4.3mm) cracks appeared after the sealing umbrella was loaded, and the same product was replaced to complete the operation, without causing harm to the patient.
 
Manufacturer Narrative
Trackwise id# (b)(4).Since the device is not available to be returned to us, a technical evaluation cannot be performed.Per our standard sop's, all events are tracked and trended to determine whether or not any trends develop.
 
Manufacturer Narrative
Trackwise #(b)(4).Updated sections: b4, g4, g7, h2, h6, h10.Analysis of production: (3331/213/67) the device history records review concluded that there were no ncmrs, rework, or deviations documented for the reported lot number.Based on the dhr/lhr review results, it was determined that there is no relation between the manufacturing process and the reported failure.Historical data analysis: (4109/213/67) the review of the historical data indicates that no other similar complaints was reported for the same lot number and reported failure mode.Trend analysis: (4110/213/67) the overall 24 month product complaint trend data for the period (b)(6) 2020 through (b)(6) 2022 was reviewed.There were no triggers identified for the review period.Communication/interviews: (4111/213/67) communication/interviews were performed to obtain all possible information.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HST III SYSTEM (4.3MM)
Type of Device
CLAMP, VASCULAR
Manufacturer (Section D)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ
Manufacturer (Section G)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ
Manufacturer Contact
arelean guzman
45 barbour pond drive
wayne, NJ 
MDR Report Key13387366
MDR Text Key286357736
Report Number2242352-2022-00096
Device Sequence Number1
Product Code DXC
UDI-Device Identifier00607567700321
UDI-Public00607567700321
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K130382
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/28/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/18/2022
Device Model NumberHST III SYSTEM (4.3MM)
Device Catalogue NumberC-HSK-3043
Device Lot Number25160806
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received03/08/2022
Was Device Evaluated by Manufacturer? No
Date Device Manufactured10/18/2021
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Age66 YR
Patient SexFemale
Patient Weight65 KG
-
-