• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SISTEMAS MEDICOS ALARIS, S.A. DE C.V. BD ALARIS PUMP MODULE SMARTSITE INFUSION SET; INTRAVASCULAR ADMINISTRATION SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SISTEMAS MEDICOS ALARIS, S.A. DE C.V. BD ALARIS PUMP MODULE SMARTSITE INFUSION SET; INTRAVASCULAR ADMINISTRATION SET Back to Search Results
Model Number 2434-0007
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/04/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported that the bd alaris pump module smartsite infusion set experienced leakage.The following information was provided by the initial reporter: while priming iv tubing for a line change, tpn started squirting out of the filter of the iv tubing.
 
Event or Problem Description
It was reported that the bd alaris pump module smartsite infusion set experienced leakage.The following information was provided by the initial reporter: while priming iv tubing for a line change, tpn started squirting out of the filter of the iv tubing.
 
Additional Manufacturer Narrative
H.6.Investigation: no product or photo was returned by the customer.The customer complaint of tpn leak from the filter could not be verified due to the product not being returned for failure investigation.A device history record review for model 2434-0007 lot number 20086789 was performed.There were no quality notifications issued for the failure mode reported by the customer during the production build of this set.The root cause of this failure could not be identified without a failure investigation.A complaint history check was performed and this is the 1st related complaint reported with the defect/condition of leakage with lot #20086789 regarding item #2434-0007.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD ALARIS PUMP MODULE SMARTSITE INFUSION SET
Common Device Name
INTRAVASCULAR ADMINISTRATION SET
Manufacturer (Section D)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer (Section G)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key13387411
Report Number9616066-2022-00038
Device Sequence Number1843416
Product Code FPA
UDI-Device Identifier37613203019690
UDI-Public37613203019690
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K944320
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date08/15/2023
Device Model Number2434-0007
Device Catalogue Number2434-0007
Device Lot Number20086789
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/06/2021
Supplement Date Received by Manufacturer02/23/2022
Initial Report FDA Received Date01/28/2022
Supplement Report FDA Received Date03/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/13/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-