No product was received as the involved device currently remains in situ and no product fault was alleged.No radiographs or testing data was provided confirming the issues the patient was experiencing with their c-reactive protein levels.All nuvasive is sterilized at the user facility prior to the case and instrument sterilization reports were not provided.Review of the reported event identified a combination of nuvasive products with a another manufactures resorbable bone regeneration material which is considered to be off label usage.The root cause of the patient infection is unknown but could be the results of patient reaction to the another manufactures resorbable bone regeneration material or an environmental contamination.No additional investigation can be completed at this time.Label review: potential adverse events and complications: as with any major surgical procedures, there are risks involved in spinal/orthopedic surgery.Infrequent.Operative and postoperative complications that may result in the need for additional surgeries include: early or late infection.Care should be used in the handling and storage of the coroent implants.The implants should not be scratched or damaged.Implants and instruments should be protected during storage and from corrosive environments.For sterile implants: assure highly aseptic surgical conditions, and use aseptic technique when removing the coroent implant from its packaging.Inspect the implant and packaging for signs of damage, including scratched or damaged devices or damage to the sterile barrier.Do not use the coroent implants if there is any evidence of damage.Refer to cleaning and sterilization instructions below for all non-sterile parts.Handling of the sterile implant - before removing the implants from the package, make sure that the protective packaging is unopened and undamaged.If the packaging is damaged, the implants have to be considered as non-sterile and may not be used.Upon removal from the package, compare the descriptions on the label with the package contents: (product number and size) note the sterile expiry date.Implants with elapsed sterile expiry dates have to be considered as non-sterile and may not be used.Take particular care that aseptic integrity is assured during removal of the implant from the inner packaging.Open the packages carefully.Take suitable measures to ensure that the implant does not come into contact with objects that could damage its surfaces.Use only the recommended instruments for implantation of the implants.Damaged implants must not be used.Sterilization - all non-sterile instruments and implants are sterilizable by steam autoclave using standard hospital practices, in addition to nuvasive¿s validated parameters.In a properly functioning and calibrated steam sterilizer, effective sterilization may be achieved using the parameters prescribed in the nuvasive cleaning nd sterilization instructions doc #(b)(4).Devices remain in-situ.
|