Model Number FV261T
Medical Device Problem Code Infusion or Flow Problem (2964)
Health Effect - Clinical Code Hydrocephalus (3272)
Date of Event 12/09/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Investigation: visual inspection: the following observations were made during the visual inspection: no abnormalities permeability test: the test showed that the shunt assistant is permeable.Computer control test: the computer controlled test showed the opening pressure of the shunt assistant, at a reference flow rate of 20 ml/h in a vertical position of the valve, to be 15.29 cmh2o.This is not within the specified tolerance of 20 cmh2o +4/-2 cmh2o.An applied pressure of 20 cmh2o, with the device in the horizontal position is expected to have a resultant pressure of 20 cmh2o +4/-2 cmh2o.Additional testing did not significantly change the results.According to our results, we determine a presence of an accelerated outflow.Internal inspection of product: in order to verify whether the investigated valve was compromised by the known risks of hydrocephalus therapy, e.G.By a build-up of natural substances (protein, blood, or tissue particles) in the cerebrospinal fluid, we have dismantled the valve.After dismantling of the valve, some deposits were found in shunt assistant.Results: based on our investigation results, we can identify an accelerated outflow at the valve.We are assuming that the deposits detected within the valve have caused the functional impairment.Deposits caused by natural substances in the cerebrospinal fluid, such as protein, blood or tissue particles, are among the known and unavoidable risks and side effects of hydrocephalus therapy.Small amounts of non-visible deposits/proteins might compromise the integrity of the valve.We can exclude a defect at the time of release.The valve met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.From our point of view, no further regulatory actions are required.
 
Event or Problem Description
It was reported that a shunt assistant (part # (b)(4)) was implanted during a procedure performed on (b)(6) 2020.According to the complainant, the shunt system was believed to be operated in underdrainage.The patient underwent a revision procedure on (b)(6) 2021.The complaint device was returned to the manufacturer for evaluation.No patient complications were reported as a result of the revision procedure.Age: (b)(6), height: 85 centimeters (cm), weight: (b)(6) (kg), gender: unknown.Same event as medwatch #: 3004721439-2021-00361.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHUNTASSISTANT 20 W/ATT DISTAL CATH
Common Device Name
HYDROCEPHALUS MANAGEMENT
Manufacturer (Section D)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM  14469
Manufacturer (Section G)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM   14469
Manufacturer Contact
joerg knebel
ulanenweg 2
potsdam, 14469
GM   14469
MDR Report Key13388279
Report Number3004721439-2021-00363
Device Sequence Number19286991
Product Code JXG
UDI-Device Identifier04041906114206
UDI-Public4041906114206
Combination Product (Y/N)N
Initial Reporter CountryRS
PMA/510(K) Number
K110206
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 01/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/21/2025
Device Model NumberFV261T
Device Catalogue NumberFV261T
Device Lot Number20045700
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/03/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/03/2022
Initial Report FDA Received Date01/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/23/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age26 MO
Patient SexUnknown
Patient Weight11 KG
-
-