• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON, DICKINSON & CO. (SPARKS) BD¿ UNIVERSAL VIRAL TRANSPORT KIT, 3 ML, FLOCKED FLEXIBLE MINITIP; CULTURE MEDIA, NON-PROPAGATING TRANSPORT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON, DICKINSON & CO. (SPARKS) BD¿ UNIVERSAL VIRAL TRANSPORT KIT, 3 ML, FLOCKED FLEXIBLE MINITIP; CULTURE MEDIA, NON-PROPAGATING TRANSPORT Back to Search Results
Model Number 220531
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 01/14/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Medical device expiration date: unknown.A device evaluation and/or device history review is anticipated, but is not complete.Upon completion, a supplemental report will be filed.Device manufacture date: unknown.
 
Event or Problem Description
It has been reported that the bd¿ universal viral transport kit, 3 ml, flocked flexible minitip has been found splashing onto the patient.The following has been provided by the initial reporter: uvt 220531 reagent splashed on a patient.Contents of the vial were spilled on customer's clothes and partially into her mouth.
 
Event or Problem Description
It has been reported that the bd¿ universal viral transport kit, 3 ml, flocked flexible minitip has been found splashing onto the patient.The following has been provided by the initial reporter: uvt 220531 reagent splashed on a patient.Contents of the vial were spilled on customer's clothes and partially into her mouth.
 
Additional Manufacturer Narrative
H.6.Investigation: this statement serves to summarize findings on the recent complaint on product 220531 (kit uvt 3ml with flocked flex minitip), lot number unknown, where it was observed that the media splashed out of the tube and onto a patient.Event description: " it was reported that uvt 220531 reagent splashed on a patient.¿ complaint history review: a review of past complaints on this product over the past 12 months does not indicate a trend on this issue.Device history record review: a review of the device history record could not be performed as a lot number was not provided.Sample analysis: no photos or returns were available.The retention samples could not be inspected as a lot number was not provided.Evaluations results: based on the investigation, no defect was observed.There is no systemic failure in the manufacturing process.As no deviations were observed in the investigation, no corrective or preventive actions are indicated at this time.Investigation conclusion: based on the evaluation of the investigation, the complaint was not confirmed.No further actions will be taken as no confirmed trend has been identified.Bd will continue to monitor for trending.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD¿ UNIVERSAL VIRAL TRANSPORT KIT, 3 ML, FLOCKED FLEXIBLE MINITIP
Common Device Name
CULTURE MEDIA, NON-PROPAGATING TRANSPORT
Manufacturer (Section D)
BECTON, DICKINSON & CO. (SPARKS)
7 loveton circle
sparks MD 21152
Manufacturer (Section G)
BECTON, DICKINSON & CO. (SPARKS)
7 loveton circle
sparks MD 21152
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key13388448
Report Number1119779-2022-00148
Device Sequence Number17377230
Product Code JSM
UDI-Device Identifier30382902205310
UDI-Public30382902205310
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K042970
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number220531
Device Catalogue Number220531
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/14/2022
Supplement Date Received by Manufacturer06/02/2022
Initial Report FDA Received Date01/28/2022
Supplement Report FDA Received Date06/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-