Model Number BD710FJ282RTS
Medical Device Problem Code Tear, Rip or Hole in Device Packaging (2385)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/05/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a patient underwent an atrial flutter right (r-afl) ablation procedure with a webster ® cs catheter with ez steer¿ technology and an open pouch seal issue occurred.It was reported that when opening the catheter for the first time the inside wrapping was opened already and the physician did not feel like the catheter was fully sterile.The catheter was replaced and the issue was resolved.The case continued.There was no patient consequence reported.Additional information was received on the event.The packaging was open.Almost as if there was no adhesive on the outer packaging.Meaning it was not fully sealed and therefore, the physician did not feel it was sterile.There was no issue with the way it was seated in the plastic.The event was assessed as a mdr reportable open pouch seal issue.
 
Additional Manufacturer Narrative
The product has not returned for analysis, however, a picture was provided by the customer.Evaluation is still in progress.The analysis has begun but is not completed at this time.When the investigational analysis has been completed, a supplemental 3500a report will be submitted.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4).
 
Additional Manufacturer Narrative
It was reported that a patient underwent an atrial flutter right (r-afl) ablation procedure with a webster ® cs catheter with ez steer¿ technology and an open pouch seal issue occurred.It was reported that when opening the catheter for the first time the inside wrapping was opened already and the physician did not feel like the catheter was fully sterile.The catheter was replaced and the issue was resolved.The case continued.There was no patient consequence reported.The investigation of the picture analysis was completed on 03-mar-2022.According to pictures provided by customer, an open pouch was observed.A manufacturing record evaluation was performed, and no internal actions related to the reported complaint condition were identified.The customer complaint was confirmed based on the picture received.However, the device has not been returned for evaluation.Since no device has been received, no product investigation can be performed.If the device is received in the future, the product analysis will be performed as appropriate in order to find the root cause of the complaint.H6.Investigation findings code of ¿appropriate term/code not available¿ represents photo/video analysis.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WEBSTER ® CS CATHETER WITH EZ STEER¿ TECHNOLOGY
Common Device Name
ELECTRODE RECORDING CATHETER OR ELECTRODE RECORDING PROBE
Manufacturer (Section D)
BIOSENSE WEBSTER INC
31 technology drive, suite 200
irvine CA 92618
Manufacturer (Section G)
BIOSENSE WEBSTER INC (JUAREZ)
circuito interior norte
1820parque industrial salvacar
juarez
MX  
Manufacturer Contact
gabriel alfageme
31 technology dr
irvine, CA 92618
9497898687
MDR Report Key13388654
Report Number2029046-2022-00172
Device Sequence Number12011674
Product Code DRF
UDI-Device Identifier10846835002430
UDI-Public10846835002430
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K090898
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/16/2024
Device Model NumberBD710FJ282RTS
Device Catalogue NumberBD710FJ282RTS
Device Lot Number30686208M
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/05/2022
Supplement Date Received by Manufacturer03/03/2022
Initial Report FDA Received Date01/28/2022
Supplement Report FDA Received Date03/12/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/17/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNKNOWN BRAND CATHETER.
Patient SexUnknown
-
-