MAUDE Adverse Event Report: CHRSITOPH MIETHKE GMBH & CO KG M.BLUE 5 SYS W/PED.CONTROL RESERVOIR; HYDROCEPHALUS MANAGEMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CHRSITOPH MIETHKE GMBH & CO KG M.BLUE 5 SYS W/PED.CONTROL RESERVOIR; HYDROCEPHALUS MANAGEMENT Back to Search Results
Model Number FX816T
Medical Device Problem Code Infusion or Flow Problem (2964)
Health Effect - Clinical Code Hydrocephalus (3272)
Date of Event 12/15/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a m.Blue (part # fx816t) was implanted during a procedure performed on (b)(6) 2021.According to the complainant, the shunt system was believed to be blocked.The patient underwent a revision procedure on (b)(6) 2021.The complaint device was returned to the manufacturer for evaluation.No patient complications were reported as a result of the revision procedure.Age: (b)(6).Height: unknown.Weight: unknown.Gender: female.
 
Additional Manufacturer Narrative
Investigation: visual inspection: the visual inspection demonstrated the following: deposits in the ped.Controlreservoir.Tissue residues and deposits on the product.Many particles in the delivery liquid.Permeability test: a permeability test has indicated that the valve has a blockage.Adjustment test: the m.Blue valve was tested and is adjustable to all specified pressures.Braking force and brake function test: the brake functionality test has shown that the brake function is fully operational and the braking force is within the given tolerances.Computer controlled test: to investigate the claim of under-drainage, the opening pressure is measured using a miethke computer controlled testing apparatus which simulates a cerebrospinal fluid flow.Because the valve is not permeable, a computer controlled test is not applicable.Internal inspection: after dismantling of the valve, some deposits were found in m.Blue.Results: based on our investigation, we confirm the presence of occlusion in the valve, likely due to the deposits observed inside the valve.As described in scientific literature, the problem encountered is one of the known, inevitable risks of hc-therapy by shunt implants.However, it is possible that the deposits observed inside the valve could have caused the malfunction in the past.As described in scientific literature, the problem encountered is one of the known, inevitable risks of hc-therapy by shunt implants.We can exclude a defect at the time of release.The valve met all specifications of the final inspection when released from christoph miethke (b)(4).No further regulatory actions are required from our point of view.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
M.BLUE 5 SYS W/PED.CONTROL RESERVOIR
Common Device Name
HYDROCEPHALUS MANAGEMENT
Manufacturer (Section D)
CHRSITOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM  14469
Manufacturer (Section G)
CHRSITOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM   14469
Manufacturer Contact
joerg knebel
ulanenweg 2
potsdam, 14469
GM   14469
MDR Report Key13388678
Report Number3004721439-2022-00001
Device Sequence Number14702268
Product Code JXG
UDI-Device Identifier04041906504526
UDI-Public4041906504526
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K192266
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 01/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/20/2026
Device Model NumberFX816T
Device Catalogue NumberFX816T
Device Lot Number20053782
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/04/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/04/2022
Initial Report FDA Received Date01/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/25/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age4 MO
Patient SexFemale
-
-