• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NEUROSURGERY TIMESH; SCREW, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NEUROSURGERY TIMESH; SCREW, FIXATION, BONE Back to Search Results
Model Number 9001640
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that during the operation, the screw slipped and broke as well as the connecting piece and cap hole broke on the plate.The products were replaced with the same product.This event resulted in a procedure delay of 1 minute.
 
Additional Manufacturer Narrative
H3) the screw was returned for analysis.There were no signs of damage on any of the returned screws.The screw head and tip were not found to be damaged.2 screws had presence of blood around the head.The laboratory personnel could not observe the same failure that occurred in the field.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TIMESH
Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
MEDTRONIC NEUROSURGERY
5290 california ave
irvine CA 92617
Manufacturer (Section G)
MEDTRONIC NEUROSURGERY
5290 california ave
irvine CA 92617
Manufacturer Contact
glen belmer
9775 toledo way
irvine, CA 92618
6122713209
MDR Report Key13389210
Report Number2021898-2022-00048
Device Sequence Number4998956
Product Code HWC
UDI-Device Identifier00613994106810
UDI-Public00613994106810
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K974017
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number9001640
Device Catalogue Number9001640
Device Lot NumberT217575
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/25/2022
Supplement Date Received by Manufacturer04/05/2022
Initial Report FDA Received Date01/28/2022
Supplement Report FDA Received Date04/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/09/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-