| Model Number |
PS TIBIAL INSERTS SZ 4, 11MM |
| Medical Device Problem Code |
Loosening of Implant Not Related to Bone-Ingrowth (4002)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
09/14/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
As reported, approximately 20 years post initial left tha, this male patient underwent revision due to aseptic loosening.No additional information is available or forthcoming.The device is not available for evaluation.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information, including the product investigation, will be submitted within 30 days of receipt.
|
| |
|
Additional Manufacturer Narrative
|
|
After further review of additional information received.(h3) the revision reported was likely the result of an insufficient bond between the implant and the bone, which led to aseptic (non-infected) loosening.The loosening was likely related to patient-related conditions, progression of degenerative joint disease, and/or longevity of the implanted devices.However, this cannot be confirmed as the devices were not available for evaluation and pre-revision x-rays were not provided.The most probable root cause associated with the reported event of "loosening¿ is associated with weakened integration of the device at the bone-implant interface due to loss of fibrous and/or bony tissue and leading to compromised anchorage of the device.Section d2a: product code.Section d2b: common device name.
|
| |
|