U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. PS TIBIAL INSERTS SZ 4, 11MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. PS TIBIAL INSERTS SZ 4, 11MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED Back to Search Results
Model Number PS TIBIAL INSERTS SZ 4, 11MM
Medical Device Problem Code Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 09/14/2020
Type of Reportable Event Serious Injury
Event or Problem Description
As reported, approximately 20 years post initial left tha, this male patient underwent revision due to aseptic loosening.No additional information is available or forthcoming.The device is not available for evaluation.
 
Additional Manufacturer Narrative
Additional information, including the product investigation, will be submitted within 30 days of receipt.
 
Additional Manufacturer Narrative
After further review of additional information received.(h3) the revision reported was likely the result of an insufficient bond between the implant and the bone, which led to aseptic (non-infected) loosening.The loosening was likely related to patient-related conditions, progression of degenerative joint disease, and/or longevity of the implanted devices.However, this cannot be confirmed as the devices were not available for evaluation and pre-revision x-rays were not provided.The most probable root cause associated with the reported event of "loosening¿ is associated with weakened integration of the device at the bone-implant interface due to loss of fibrous and/or bony tissue and leading to compromised anchorage of the device.Section d2a: product code.Section d2b: common device name.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PS TIBIAL INSERTS SZ 4, 11MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer Contact
kate jacobson
2320 nw 66 court
gainesville, FL 32653
3523771140
MDR Report Key13389353
Report Number1038671-2022-00066
Device Sequence Number1715754
Product Code JWH
UDI-Device Identifier10885862048394
UDI-Public10885862048394
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K933610
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2000
Device Explanted Year2020
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 07/23/2022
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/19/2015
Device Model NumberPS TIBIAL INSERTS SZ 4, 11MM
Device Catalogue Number204-24-11
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/07/2022
Supplement Date Received by Manufacturer06/07/2022
Initial Report FDA Received Date01/28/2022
Supplement Report FDA Received Date07/23/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/21/2011
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-