Model Number OPTI0407CSF10
Medical Device Problem Code Inaccurate Information (4051)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/27/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that: "they say inside de external the reference was different than the one on the external box outside card box label opti0407csf10 f210500135, inside contain opti0204csf10 du lot f210500134.Product was not use and pouch remain un opened (closed) we can bring back all before use.".
 
Additional Manufacturer Narrative
Balt usa reference complaint #(b)(4).The unit with the label discrepancy has been returned to balt usa under rga (b)(4).This complaint has triggered the opening of hhe-(b)(4) (balt usa co#(b)(4)) and ncmr (b)(4) to drive completion of the hhe.Any further corrective action and investigation will be documented under the hhe and ncmr.(b)(6) 2021: while preparing hhe-(b)(4), the initial assessment of this complaint file was re-evaluated due to feedback received from the regulatory department regarding the hhe.Upon initial assessment, the assumption of the label discrepancy was made that the shelf carton label was incorrect, and the pouch label was correct according to the lhr.As recorded in balt usa ncmr (b)(4), one additional unit was returned from balt extrusion under rga (b)(4) ((b)(6) 2021) after this complaint had already been closed.At the time of the initial assessment, the device pouch was not opened knowing that the unit had potential to be reworked.During the re-evaluation of this complaint, the pouch of both units were opened and the length of the implant was found to be 7cm, therefore the shelf carton label is correct and the pouch label is incorrect.This additional evaluation has been documented in balt usa ncmr (b)(4) and hhe-(b)(4).(b)(6) 2022: initial reportability decision tree deemed this event as not reportable.Reportability decision tree was not updated after the additional information was received regarding this incident.Reportability decision tree has been updated and has been deemed reportable.Notification of urgent field safety notice (fsn) / product recall was issued 04jan2022 with the reference "fsn #: 3014162263-01/04/22-001-r" for inside the united states.Additional reference "(b)(4)_optima labeling mix-up" used for outside the united states.Fda acknowledgement letter reference res# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTIMA COIL SYSTEM
Common Device Name
OPTIMA
Manufacturer (Section D)
BALT USA
29 parker
irvine CA 92618
Manufacturer Contact
moises colin
29 parker
irvine, CA 92618
9497881443
MDR Report Key13389359
Report Number3014162263-2022-00005
Device Sequence Number13881973
Product Code HCG
UDI-Device Identifier00818053026096
UDI-Public00818053026096
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K172390
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial
Report Date (Section B) 01/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date05/15/2026
Device Model NumberOPTI0407CSF10
Device Catalogue NumberN/A
Device Lot NumberF210500135
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/24/2021
Type of Report(Section G)Five-Day
Initial Date Received by Manufacturer 08/27/2021
Initial Report FDA Received Date01/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/15/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-