• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AMO PUERTO RICO MFG. INC. TECNIS CL IOL; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMO PUERTO RICO MFG. INC. TECNIS CL IOL; INTRAOCULAR LENS Back to Search Results
Model Number Z9002
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Eye Injury (1845)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Ethnicity/race: unknown, information not provided.Date of event: unknown, not provided, but the best estimate date is between oct 20, 2021 and dec 28, 2021.The intraocular lens (iol) has not returned for analysis.Therefore, a failure analysis of the complaint device cannot be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.Attempts have been made to obtain additional information regarding the event; however, the information has not been received.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that an intraocular lens (iol) was explanted from a patient's right eye in a secondary surgical procedure.A replacement lens of the same model but different diopter (17.5) was used.No further information was provided.
 
Additional Manufacturer Narrative
Additional information: section d9: device available for evaluation? yes.Section d9: returned to manufacturer: yes.Section d9: date returned to manufacturer: feb 1, 2022.Section h3: device evaluated by manufacturer: yes.Device evaluation: visual inspection under magnification revealed viscoelastic residue on the optic body and haptics and that the lens was received cut in half with detached haptics, which is consistent with a lens that was handled during explant.Based on the return condition of the lens no further product evaluation could be performed.The complaint issue could not be confirmed, and no product deficiency could be identified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS CL IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key13389364
Report Number3012236936-2022-00253
Device Sequence Number1817483
Product Code HQL
UDI-Device Identifier05050474530683
UDI-Public(01)05050474530683(17)260817
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P880081
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/17/2026
Device Model NumberZ9002
Device Catalogue NumberZ900200195
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/03/2022
Supplement Date Received by Manufacturer03/04/2022
Initial Report FDA Received Date01/28/2022
Supplement Report FDA Received Date03/30/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/17/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age72 YR
Patient SexFemale
Patient Weight65 KG
-
-