PHYSIO-CONTROL, INC. - 3015876 LIFEPAK(R) 15 MONITOR/DEFIBRILLATOR; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE) Back to Search Results
Model Number 15
Medical Device Problem Code Electrical /Electronic Property Problem (1198)
Health Effect - Clinical Code Ventricular Fibrillation (2130)
Date of Event 12/31/2021
Type of Reportable Event Malfunction
Event or Problem Description
The customer contacted stryker to report that their device was unable to deliver defibrillation therapy during a patient event.The customer advised that the device was charged to 100j, when "synch" was pressed the monitor flickered, then the screen turned black.After a few seconds the screen returned.The device was charged to 100j, then dumped, and charged to 360j when monitor flickered again and power cycled.The service indicator was then illuminated.In addition the device logged a potentially critical event code that is indicative of a device lockup.This issue is patient related; however the patient was saved by the medics.
 
Additional Manufacturer Narrative
Stryker contacted the customer to request additional information on the patient.The customer provided stryker with all the available patient information.Stryker evaluated the customer¿s device but was unable to duplicate the reported issue.The device software was reloaded as a precaution.After other unrelated repairs were completed, proper device operation was observed through functional and performance testing.The device was returned to the customer for use.The cause of the reported issue could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK(R) 15 MONITOR/DEFIBRILLATOR
Common Device Name
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer (Section G)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer Contact
todd bandy
11811 willows road ne
redmond, WA 98052
4258674000
MDR Report Key13389380
Report Number0003015876-2022-00122
Device Sequence Number1738169
Product Code MKJ
UDI-Device Identifier00883873912799
UDI-Public00883873912799
Combination Product (Y/N)N
Initial Reporter StateDE
Initial Reporter CountryUS
PMA/510(K) Number
K142430
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 01/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number15
Device Catalogue Number99577-001217
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/19/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/05/2022
Initial Report FDA Received Date01/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/02/2014
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient Age61 YR
Patient SexMale
Patient Weight100 KG
Patient RaceWhite
-
-