• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MASIMO - 15750 ALTON PKWY RD SET INF CS-2; OXIMETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MASIMO - 15750 ALTON PKWY RD SET INF CS-2; OXIMETER Back to Search Results
Model Number 4472
Medical Device Problem Code Incorrect Measurement (1383)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/11/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional manufacuring narrative: attempts have been made to obtain the product.The product has not been returned to masimo to allow an analysis to be performed.If the product is returned for evaluation or new information is obtained, a follow up report will be submitted., other, other text: initial reporter zip code exceeded the maximum allowable characters, zip code is as follows: (b)(6).
 
Event or Problem Description
The customer reported the device was inaccurate.No consequences or impact to patient were reported.
 
Event or Problem Description
The customer reported the device was inaccurate.No consequences or impact to patient were reported.
 
Additional Manufacturer Narrative
Additional manufacturing narrative: the returned sensor was evaluated.During evaluation the sensor passed all visual and functional testing.  during simulation testing, the sensor passed manual and preset conditions and provided accurate measurements.The sensor was determined to be functioning as designed.  , other, other text: additional information: d4 udi# and expiration date; h4 manufacturing date initial reporter zip code exceeded the maximum allowable characters, zip code is as follows: 3015 gd, corrected data: d1 brand name; d4 model, lot #, catalog #; and g5 510(k) #.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RD SET INF CS-2
Common Device Name
OXIMETER
Manufacturer (Section D)
MASIMO - 15750 ALTON PKWY
15750 alton pkwy
irvine CA 92618
Manufacturer (Section G)
MASIMO - MEXICALI
industrial vallera de mexicali calzada del oro, no.2001
mexicali, baja california 21600
MX   21600
Manufacturer Contact
matofa fautua
15750 alton pkwy
irvine, CA 92618
9498223661
MDR Report Key13389533
Report Number3019388613-2022-00029
Device Sequence Number4858083
Product Code DQA
UDI-Device Identifier10843997013212
UDI-Public10843997013212
Combination Product (Y/N)N
Initial Reporter CountryNL
PMA/510(K) Number
K042536
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date (Section B) 01/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/01/2024
Device Model Number4472
Device Catalogue Number4472
Device Lot Number21JZB
Was Device Available for Evaluation? Device Returned to Manufacturer
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/04/2022
Supplement Date Received by Manufacturer03/01/2022
Initial Report FDA Received Date01/28/2022
Supplement Report FDA Received Date03/25/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-